Drug Abuse Test Kit K2 (Synthetic Marijuana) Rapid Diagnostic Test strip, professional use, urine drug test, 50ng/ml
Product Name
K2 (Synthetic Marijuana) Rapid Diagnostic Test Strip
Specimen: Urine
Product | Cat No | Specimen | Sensitivity | Format | Kit Size |
K2 | YD131S3 | Urine | 50 ng/mL | 3mm Strip | 100T |
YD131C4 | 4mm Cassette | 25T |
Intended Use
The Synthetic Marijuana (K2) Rapid Test Strip (Urine) is a rapid chromatographic immunoassay for the detection of Synthetic Marijuana metabolite in human urine. The synthetic marijuana detected by the test includes, but are not limited to, the metabolites of JWH-018 and JWH-073. This assay provides only a preliminary analytical test result. A more specific alternate chemical method must be used in order to obtain a confirmed analytical result. Gas chromatography/mass spectrometry (GC/MS) is the preferred confirmatory method. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are used.
Parameter | Calibrator | Cut-off (ng/mL) |
K2(Synthetic Marijuana) | Synthetic Marijuana | 50 |
Test Principle
The K2 Rapid Test Cassette (Urine) is an immunoassay based on the principle of competitive binding. Drugs which may be present in the urine specimen compete against the drug conjugate for binding sites on the antibody. During testing, a urine specimen migrates upward by capillary action. Synthetic Marijuana metabolite, if present in the urine specimen below 50ng/mL, will not saturate the binding sites of antibody-coated particles in the test device. The antibody-coated particles will then be captured by immobilized synthetic marijuana conjugate and a visible colored line will show up in the test line region. The colored line will not form in the test line region if the Synthetic Marijuana metabolite level exceeds 50ng/mL because it will saturate all the binding sites of anti- Synthetic Marijuana antibodies. A drug-positive urine specimen will not generate a colored line in the test line region, while a drug-negative urine specimen or a specimen containing a drug concentration less than the cut-off will generate a line in the test line region. To serve as a procedural control, a colored line will always appear at the control line region indicating that proper volume of specimen has been added and membrane wicking has occurred.
DIRECTIONS FOR USE
INTERPRETATION OF RESULTS
NEGATIVE:* A colored line appears in the Control region (C) and colored lines appears in the Test region (T). This negative result means that the concentrations in the urine sample are below the designated cut-off levels for a particular drug tested.
*NOTE: The shade of the colored lines(s) in the Test region (T) may vary. The result should be considered negative whenever there is even a faint line.
POSITIVE: A colored line appears in the Control region (C) and NO line appears in the Test region (T). The positive result means that the drug concentration in the urine sample is greater than the designated cut-off for a specific drug.
INVALID: No line appears in the Control region (C). Insufficient specimen volume or incorrect procedural techniques are the most likely reasons for Control line failure. Read the directions again and repeat the test with a new test card. If the result is still invalid, contact your manufacturer.
PERFORMANCE:
Accuracy
A side-by-side comparison was conducted using the K2 Rapid Test Dipstick (Urine) and GC/MS. The following results were tabulated:
Method | GC/MS | Total Results | ||||
K2 Rapid | Results | Positive | Negative | |||
Positive | 78 | 3 | 81 | |||
Test Dipstick | ||||||
Negative | 2 | 167 | 169 | |||
Total Results | 80 | 170 | 250 | |||
% Agreement | 97.5% | 98.2% | 98.0% |
Analytical Sensitivity
A drug-free urine pool was spiked with K2 at the following concentrations: 0,25,37.5,50,62.5,75 and 150ng/mL. The results demonstrate >99% accuracy at 50% above and 50% below the cut-off concentration. The data are summarized below:
Synthetic Marijuana | Percent of | n | Visual Result | |
Concentration (ng/mL) | Cut-off | Negative | Positive | |
0 | 0% | 30 | 30 | 0 |
25 | -50% | 30 | 30 | 0 |
37.5 | -25% | 30 | 26 | 4 |
50 | Cut-off | 30 | 15 | 15 |
62.5 | +25% | 30 | 3 | 27 |
75 | +50% | 30 | 0 | 30 |
150 | 3X | 30 | 0 | 30 |