Add to Cart
Drug Abuse Test Kit Acetaminophen Rapid Test 3mm Strip ,high sensitivity cut-off 400ng/ml
Product Name
ACE (Acetaminophen) Rapid Diagnostic Test
Specimen: Urine
Background
Acetaminophen is one of the most commonly used drugs, yet it is also an important cause of serious liver injury. Acetaminophen is the generic name of a drug found in many common brand name over-the-counter (OTC) products, such as Tylenol, and Prescription (Rx) products, such as Vicodin and Percocet. Acetaminophen is an important drug, and its effectiveness in relieving pain and fever is widely known. Unlike other commonly used drugs to reduce pain and fever (e.g., nonsteroidal antinflammatory drugs (NSAIDs), such as aspirin, ibuprofen, and naproxen), at recommended doses acetaminophen does not cause adverse effects, such as stomach discomfort and bleeding, and acetaminophen is considered safe when used according to the directions on its OTC or Rx labeling. However, taking more than the recommended amount can cause liver damage, ranging from abnormalities in liver function blood tests, to acute liver failure, and even death. Many cases of overdose are caused by patients inadvertently taking more than the recommended dose (i.e., 4 grams a day) of a particular product, or by taking more than one product containing acetaminophen (e.g., an OTC product and an Rx drug containing acetaminophen). The mechanism of liver injury is not related to acetaminophen itself, but to the production of a toxic metabolite. The toxic metabolite binds with liver proteins, which cause cellular injury. The ability of the liver to remove this metabolite before it binds to liver protein influences the extent of liver injury. The ACE Rapid Test Cassette (Urine) is a rapid urine screening test that can be performed without the use of an instrument. The test utilizes a monoclonal antibody to selectively detect elevated levels of Acetaminophen in urine. The ACE Rapid Test Cassette (Urine) yields a positive result when Acetaminophen in urine exceeds 5,000ng/mL.
Test Principle
The ACE Rapid Test Cassette (Urine) is an immunoassay based on the principle of competitive binding. Drugs which may be present in the urine specimen compete against the drug conjugate for binding sites on the antibody. During testing, a urine specimen migrates upward by capillary action. Acetaminophen, if present in the urine specimen below 5,000ng/mL, will not saturate the binding sites of antibody-coated particles in the test. The antibody-coated particles will then be captured by immobilized Acetaminophen conjugate and a visible colored line will show up in the test line region. The colored line will not form in the test line region if the Acetaminophen level exceeds 5,000ng/mL because it will saturate all the binding sites of anti-Acetaminophen antibodies. A drug-positive urine specimen will not generate a colored line in the test line region because of drug competition, while a drug-negative urine specimen or a specimen containing a drug concentration lower than the cut-off will generate a line in the test line region. To serve as a procedural control, a colored line will always appear in the control line region, indicating that proper volume of specimen has been added and membrane wicking has occurred
SPECIMEN COLLECTION AND STORAGE
The urine specimen must be collected in a clean and dry container. Urine collected at any time of the day may be used. Urine specimens exhibiting visible particles should be centrifuged, filtered, or allowed to settle to obtain clear specimen for testing. Specimen.
Urine specimens may be stored at 2-8°C for up to 48 hours prior to testing. For long-term storage, specimens may be frozen and stored below -20°C. Frozen specimens should be thawed and mixed before testing.
STORAGE
Store as packaged in the sealed pouch either at room temperature or refrigerated (2-30°C). The test is stable through the expiration date printed on the sealed pouch. The test must remain in the sealed pouch until use. DO NOT FREEZE. Do not use beyond the expiration date.
LIMITATION
1. The 6-MAM Rapid Test Cassette provides only a qualitative, preliminary analytical result. A secondary analytical method must be used to obtain a confirmed result. Gas chromatography/mass spectrophotometry (GC/MS) is the preferred confirmatory method2.3.
2. It is possible that technical or procedural errors, as well as other interfering substances in the urine specimen may cause erroneous results.
3. Adulterants, such as bleach and/or alum, in urine specimens may produce erroneous results regardless of the analytical method used. If adulteration is suspected, the test should be repeated with another urine specimen.
4. A positive result indicates presence of the drug or its metabolites but does not indicate level of intoxication, administration route or concentration in urine.
5. A negative result may not necessarily indicate drug-free urine. Negative results can be obtained when drug is present but below the cut-off level of the test.
Performance characteristics
A side-by-side comparison was conducted using the Rapid Test Cassette (Urine) and GC/MS. The following results were tabulated:
Table 1: Specimen Correlation
Method | GC/MS | Total Results | ||||
ACE Rapid Test | Results | Positive | Negative | |||
Positive | 23 | 2 | 25 | |||
Cassette (Urine) | ||||||
Negative | 2 | 68 | 70 | |||
Total Results | 25 | 70 | 95 | |||
% Agreement | 92.0% | 97.1% | 95.8% |
ORIENT NEW LIFE MEDICAL CO., LTD. | |
Contact: | Jerry Meng |
Email: | Jerry @ newlifebiotest .com |
Tel. | +86 18657312116 |
SKYPE | enetjerry |